Genizer™ located in technology park of Irvine in great Los Angeles, is dedicated to advancing homogenizer nanotechnology.
Temperature-Pressure Dual-Parameter Transmitter,Sanitary Gauges & Sensors. High Pressure Homogenizer. Troubleshooting: Our design has minimized the occurrence of the air-blocking, chamber-blocking and overload.
However, the improper operation of the homogenizer can still induce the error and damage of the homogenizer. Air Blocking Symptom: There is no pressure or there is no sucking from inlet Prevention: Do not run with empty inlet Reason: There is air inside the pumpSolution: 1) In the Customer setup, push the button: Degas 2) Running 50mL water with 50% speed 3) Push the Ball in the inlet valve with a pin.